PRECISE|OSCILLOSCOPE|DCA 10021

DCA 10021 _ 10G Oscilloscope

The DCA10021 optical sampling oscilloscope is used to test transmitter performance indicators. It supports optical and electrical interface inputs. The optical channel has typical filters for optical ports of 10Gbps and below, and the electrical channel bandwidth is up to 20GHz.

It adopts the best vertical resolution and high sampling rate ADC to achieve continuous equivalent sampling technology, and combines the algorithm to complete the eye pattern reconstruction. The excellent circuit design ensures that the system noise is as low as 3uW. It is paired with a calibrated reference receiver (in line with industry tolerances). Standard), perfectly meeting the industry's needs for accurate testing of digital waveforms.

Basic Specification

ItemSpecification
Channel No.Optical channel x 1
Electrical channel x 1
Filter1.229、1.244、1.25、2.5、10Gb/s
2.458、2.488、2.5Gb/s
3.072、3.125Gb/s
6.144、6.25Gb/s,
[Note] Other speeds can be customized
Temperature MonitorSupport
InterfaceLAN
[Note] Other interfaces can be customized
Max. Power30W

Features

  • High resolution and high sampling ADC.
  • Electrical channel bandwidth up to 20GHz.
  • The optical channel supports 1~11Gbps different rate filter options.
  • The optical receiver supports single-mode and multi-mode interfaces.
  • ER self-calibration, dark current calibration.
  • Supports common mask and margin tests in the industry.

Technical Specifications

ItemSpec.
Optical Channel Bandwidth10GHz
Wavelength Range750nm~ 1650nm
RMS Noise3uw
Sensitivity-12dBm
Average Power (Consistency)±0.1dB
Extinction Ratio (Conformity)±0.1dB
Template Margin (consistency)±2.5%
Max. Input
(without destroying peak power)
+3dBm
Max. Input
(Linear test range)
0dBm
Monitor Average Power Measurement Range-17dBm~ +3dBm
Input Return Loss30dB
Optical Fiber Interface62.5/125um FC/UPC
ItemSpec.
Max. Input Bandwidth20GHz
Max. Input Range1Vpp
RMS Noise< 1mV
Sensitivity< 50mV
Input Impedance50Ω
Return Loss< -18dB
ItemSpec.
Clock Input Bandwidth80MHz~4GHz
Clock Input Sensitivity<50mV
Trigger Max. Input≤±2V
Trigger Input Impedance50Ω
Time Base Interval Accuracy<64fs
ItemSpec.
Sampling Point Quantity≤3500 Point/Waveforms@10Gbps
≤9000 Point/Waveforms@1.25Gbps
Cumulative Waveform
Range
1~10 million Waveforms
[Note] Other waveform ranges can be customized
Refresh Rate100 Waveforms/S

Typical Data

Software Execution Interface

Optional Items

10G Clock Recovery Module
ItemSpec.
#1Not equipped
#2Equipped

TOP