PRECISE|Comprehensive Tester|RCML 20192

RCML 20192 _ Detector Coupling Tester

RCML20192 detector coupling tester is used for coupling production and testing of various pin-position ROSA/BOSA devices of APD and PIN-TIA (Vpd/Mon).

This instrument comes with a light attenuation module, which can achieve rapid attenuation. There is a built-in high-precision optical power meter for output monitoring, which can monitor the optical power of the output port in real time during attenuation. It also supports external optical module interfaces and can be connected to different wavelength modules to provide different coupling light source requirements for the production of ROSA/BOSA devices.

Features

  • Customized test conditions can meet various test requirements. It is compatible with two types of detectors, APD and PIN-TIA (Vpd/Mon), and supports devices with different pin packages.
  • Provides four built-in light sources of 850nm, 1310nm, 1490nm and 1550nm, which can output analog modulation frequencies in the range of 1M, 2M and 4M.
  • Built-in light attenuation module.
  • Supports SFP and QSFP assembly optical modules
  • Outputs 0~70V continuously adjustable voltage.
  • Response current measurement range 0~9mA.
  • There are protection circuits against overshoot, kickback, ESD and surge.
  • Support return loss test (optional).

Technical Specifications

ItemSpec.
Breakdown Voltage (Vbr)0~70V, Accuracy 0.1%FS±0.07V
Dark Current (Id)Measuring range:0~100nA,Accuracy 1%FS±1nA
Response Current (Iapd)Measuring range:
Range 1: 0~1000nA, Accuracy 0.1%FS±1nA
Range 2: 0~10uA, Accuracy 0.1%FS±10nA
Range 3: 0~100uA, Accuracy 0.1%FS±0.1uA
Range 4: 0~1000uA, Accuracy 0.1%FS±1uA
Range 5: 0~9000uA, Accuracy 0.1%FS±9uA
Performance indicators detected during APD coupling.
ItemSpec.
Dark Current (Id)Measuring range:0~240nA,Accuracy 1%FS±2.4nA
Response Current (Imon)Measuring range:
Range 1: 0~2400nA, Accuracy 0.1%FS±2.4nA
Range 2: 0~24uA, Accuracy 0.1%FS±24nA
Range 3: 0~240uA, Accuracy 0.1%FS±0.24uA
Range 4: 0~2400uA, Accuracy 0.1%FS±2.4uA
Range 5: 0~9000uA, Accuracy 0.1%FS±9uA
Performance indicators detected during PIN-TIA(Mon) coupling.
ItemSpec.
TIA Voltage
(Vcc-TIA)
Output voltage:0~5V, Accuracy 0.1%FS±20mV
TIA Current
(Icc)
Measuring range:0~100mA, Accuracy 0.1%FS±1mA
Signal Reception Strength
(D+、D)
Range:0~3.3V, Accuracy 1%FS±2mV
Return LossMeasuring range:10~50dB, Accuracy ±0.3db
External Light Source InterfaceSupports SFP and QSFP assembly optical modules
Calibration Wavelength1310nm、1490nm、1550nm,
[Note] Other wavelengths can be customized.
Attenuation Range0~40dB
Attenuation Accuracy≤±0.2dB

Typical Data

EOS Test
APD power-on waveform
APD power-off waveform
TIA power-on waveform
TIA power-off waveform

Optional Items

Return Loss Function
ItemSpec.
#1Not support return loss
#2Support loss recovery

TOP