PRECISE | Source Meter | PLT2301

PLT2301 High Power Pulse Source Meter

The PLT2301 pulse source is suitable for testing high-power lasers such as visible light, TO type Lidar , and VCSEL arrays. It supports pulse and DC PIV curve scanning tests, characteristic parameter calculations, and spectral characteristic tests (an external spectrometer is required).

The instrument integrates a lot of functions such as pulse voltage source, current meter, voltage meter and optical power meter, and has the characteristics of small dimension, fast testing speed and easy to use.

Features

  • Support us level pulse, DC output and test.
  • Min. pulse width is 10us and max. duty ratio up to 60%.
  • Max. voltage 40V and driving current can up to 30A.
  • Support Supports voltage 2-wire and 4-wire measurement, to maximize measurement accuracy.
  • Support multi-devices combination, to provide higher output current.
  • Supports high-speed acquisition and drawing of any single pulse waveform under continuous pulses, and real-time monitoring of pulse waveforms.
  • Supports 2-channel sync pulse trigger output, with adjustable polarity and sync pulse width.
  • Supports output short-circuit protection and over-temperature protection.
  • PD response wavelength range: 850~1700nm.

Technical Specifications

DC Performance
ItemSpec.
Output Current0~0.1A,Accuracy 0.1%RDG±0.1mA;
0~0.5A,Accuracy 0.1%RDG±0.5mA;
0~0.5A,Accuracy 0.1%RDG±0.5mA
Output Voltage10V
Current Measurement0~0.1A,Accuracy 0.1%RDG±0.1mA;
0~0.5A,Accuracy 0.1%RDG±0.5mA;
0~5A,Accuracy 0.1%RDG±5mA
Voltage Measurement0~10V,Accuracy 0.1%RDG±10mV
Pulse Performance
ItemSpec.
Output Current0~5A,Accuracy 0.1%RDG±5mA
Output VoltageGear 1: 10V
Gear 2:40V
Pulse Width10μs~5ms  

[Note] Full width at half max.
Pulse Width Resolution10ns
Pulse Rising and Falling time≤20% x Pulse Width
Pulse Width Accuracy≥20ns
Pulse Width Duty RatioOutput Voltage 10V,D≤5/I×100%;
Output Voltage 40V,D≤0.83/I×100%

[Note] D is the duty ratio, the max. doesn't exceed 60%; I is the load current.
Current Detection Range and Accuracy0~5A,Accuracy 0.1%RDG±5mA;
0~30A,Accuracy 0.1%RDG±30mA
Voltage Detection Range and Accuracy0~10V,0.1%RDG±10mV;
0~60V,0.1%RDG±60mV
Optical Port Power
ItemSpec.
Optical Port Power Detection Range and Accuracy0~100uW,Accuracy 0.1%RDG±0.1uW;
0~1mW,Accuracy 0.1%RDG±1uW;
0~5mW,Accuracy 0.1%RDG±5uW
Electrical Port Current
ItemSpec.
Electrical Port Current Detection Range and Accuracy0~100uA,Accuracy 0.1%RDG±0.1uA;
0~1mA,Accuracy 0.1%RDG±1uA;
0~10mA,Accuracy 0.1%RDG±10uA
MPD
ItemSpec.
MPD + Reverse Bias Voltage0~5V,Accuracy 1%RDG±50mV
Current Detection Range and Accuracy0~100uA,Accuracy 0.1%RDG±0.1uA;
0~1mA,Accuracy 0.1%RDG±1uA;
0~10mA,Accuracy 0.1%RDG±10uA

Measured Data Example

Test data of TO type Laser

TOP