It adopts the best vertical resolution and high sampling rate ADC to achieve continuous equivalent sampling technology, and combines the algorithm to complete the eye pattern reconstruction. The excellent circuit design ensures that the system noise is as low as 3uW. It is paired with a calibrated reference receiver (in line with industry tolerances). Standard), perfectly meeting the industry's needs for accurate testing of digital waveforms.
Basic Specification
Item | Specification |
---|---|
Channel No. | Optical channel x 1 Electrical channel x 1 |
Filter | 1.229、1.244、1.25、2.5、10Gb/s 2.458、2.488、2.5Gb/s 3.072、3.125Gb/s 6.144、6.25Gb/s, [Note] Other speeds can be customized |
Temperature Monitor | Support |
Interface | LAN [Note] Other interfaces can be customized |
Max. Power | 30W |
Features
- High resolution and high sampling ADC.
- Electrical channel bandwidth up to 20GHz.
- The optical channel supports 1~11Gbps different rate filter options.
- The optical receiver supports single-mode and multi-mode interfaces.
- ER self-calibration, dark current calibration.
- Supports common mask and margin tests in the industry.
Technical Specifications
Optical Channel Parameter
Item Spec. Optical Channel Bandwidth 10GHz Wavelength Range 750nm~ 1650nm RMS Noise 3uw Sensitivity -12dBm Average Power (Consistency) ±0.1dB Extinction Ratio (Conformity) ±0.1dB Template Margin (consistency) ±2.5% Max. Input
(without destroying peak power)+3dBm Max. Input
(Linear test range)0dBm Monitor Average Power Measurement Range -17dBm~ +3dBm Input Return Loss 30dB Optical Fiber Interface 62.5/125um FC/UPC Electrical Channel Parameter
Item Spec. Max. Input Bandwidth 20GHz Max. Input Range 1Vpp RMS Noise < 1mV Sensitivity < 50mV Input Impedance 50Ω Return Loss < -18dB Trigger Parameter
Item Spec. Clock Input Bandwidth 80MHz~4GHz Clock Input Sensitivity <50mV Trigger Max. Input ≤±2V Trigger Input Impedance 50Ω Time Base Interval Accuracy <64fs Sampling Parameter
Item Spec. Sampling Point Quantity ≤3500 Point/Waveforms@10Gbps
≤9000 Point/Waveforms@1.25GbpsCumulative Waveform
Range1~10 million Waveforms
[Note] Other waveform ranges can be customizedRefresh Rate 100 Waveforms/S
Typical Data