PRECISE|Comprehensive Tester|PIV LS

PIV Intelligent Coupled Power Supply

PIV-LS is suitable for online monitoring of LD, TOSA, and BOSA coupling. The system provides driving power for the LD in the OSA component and real-time monitoring of related performance indicators during the coupling process.

This machine can display relevant parameters, as well as P-I curve, Im-I curve, V-I curve, etc. It can also be connected to an external oscilloscope to display relevant waveforms.

Supports coupling testing of LD's common anode and individually packaged devices.

Features

  • Supports common anode and independent package mode laser coupling.
  • Real-time detection and display of threshold current, drive current, optical power, backlight current, forward voltage and other parameters.
  • Supports six coupling modes: Ith+X, Ith+X+Im, fixed-point X, fixed-point X+Im, DC X, and DC Ith+X.
  • The optical power detection adopts 3 gears to achieve the steepness of the curve at different gears.
  • The backlight current detection adopts 4 gears to achieve the steepness of the curve at different gears.
  • Support automatic gear shifting and manual gear shifting
  • Support wavelength 800~1700nm
  • Built-in high-efficiency anti-surge protection circuit and slow start circuit to achieve damage-free laser coupling.

Technical Specifications

ItemSpec.
Laser Assembly TypeCommon anode configuration and independent type
Coupling MethodIth+X (without backlight)、Ith+X (with backlight)、Fixed point X (without backlight)、Fixed point X (with backlight)、Ith+X (DC fixed point)、X (DC fixed point)
Test ItemsThreshold current、Optical power、Drive current、Backlight current、Forward voltage、Wavelength、etc.
Wavelength1310nm、1490nm、1550nm、850nm
[Note]Special wavelengths can be customized
Drive CurrentRange:0~75mA adjustable,Error <0.1mA
Backlight CurrentRange:0~2500uA,Error <15uA
Optical PowerLevel 1:0~10uW or 0~1000uW;
Level 2:0~1500uW or 0~2500uW;
Level 3:0~5000uW
Error <0.2dB
Multiple gears can be customized to test up to 10mW
Forward Voltage<2.5V,Error <30mV
Curve Refresh Rate<20ms
Calibration FunctionSupports P、V、Im multiple parameter calibration
Warning FunctionBuzzer warning for abnormal conditions, power-off protection
Power Gear SwitchingSupport automatic gear shifting and manual gear shifting
Detector Detection Wavelength800~1700nm
Optical Port Detector Diameter1.0mm / 2.0mm / 3.0mm / 3x4mm
Optical Port Four-way Difference≤0.1dB

TOP

en_USEnglish