PRECISE|Comprehensive Tester|LIV 60301

LIV 6 Series LIV Comprehensive Tester

LIV60301 comprehensive tester is suitable for LIV characteristic curve, PV characteristic curve and characteristic parameter testing of LD, TO-Can, and TOSA, EML and butterfly devices with TEC function in the optical communication industry, also can support spectral characteristic testing (an external spectrometer is required).

The system integrates the functions of temperature controller, current source, voltage source, ammeter, voltmeter and optical power meter.

Features

  • Supports pulse spectrum testing.
  • Support multi-mode TOSA in 50um and 62.5um optical fiber transmission comparison.
  • FP / DFB / VCSEL / DML / EML / SOA-EML devices in any assembly type
  • Supports LD Reverse Leakage Current Test
  • LD Reverse Vbr Test
  • Supports MPD Dark Current Test
  • Support LIV Scanning, Step time 2ms~6s optionally configured
  • Supports PV curve test of EA and DC-ER test.
  • EA Reverse Vbr Test
  • Support TEC temperature control
  • Max. Support 5 channel driving current and 1 channel driving voltage simultaneously.
  • Provides three light source input ports: built-in PIN, external PIN interface and integrating sphere interface.
  • Built-in PIN response wavelength range: 800~1700nm.

Technical Specifications

ItemSpec.
Laser Assembly TypeFP / DFB / VCSEL / DML / EML / SOA-EML devices in any assembly type
Test ItemsLIV curve,PV curve,Ith,Po,Vf,Im,Rs,SE,Kink,Vinf,DC-ER, etc.
LD Drive Current OutputRange:0~250mA,Accuracy 0.1%FS±0.25mA,max. customizable to 500mA
LD Forward Voltage TestRange:0~5.0V,Accuracy 0.1% FS±50mV
LD Reverse Leakage Current Test0~100nA,Accuracy 1% FS±1.0nA,@<50%RH
0~1uA,Accuracy 0.1% FS±10nA,@<50%RH
0~10uA,Accuracy 0.1% FS±0.1uA,@<50%RH
0~100uA,Accuracy 0.1% FS±1uA,@<50%RH
0~1mA,Accuracy 0.1% FS±10uA,@<50%RH
LD Backward Voltage OutputRange:0~30.0V,Accuracy 0.1% FS±0.3V
LD Reverse Current Output0~10uA,Accuracy 0.1% FS±0.1uA
0~100uA,Accuracy 0.1% FS±1uA
0~1mA,Accuracy 0.1% FS±10uA
LD Reverse Vbr TestRange:0~30.0V,Accuracy 0.1% FS±0.3V
MPD Backward Voltage OutputRange:0~5.0V,Accuracy 0.1% FS±50mV
MPD Dark Current TestRange:0~300nA,Accuracy 0.5% FS±3.0nA,@<50%RH
MPD Backlight Current TestRange:0~3000uA,Accuracy 0.1% FS±10uA
Four-channel Extended Current OutputRange:0~250mA,Accuracy 0.1%FS±0.25mA,max. customizable to 500mA
Four-channel Extended Current and Voltage Drop TestRange:0~5.0V,Accuracy 0.1% FS±50mV
Built-in PIN Tube Detects Optical PowerRange:0~10mW,Accuracy 0.1% FS±50uW
[Note]For higher power testing, please choose: external PIN tube or integrating sphere
Built-in PIN Tube Response Wavelength800~1700nm (Supports 850nm VCSEL device testing)
Correction FunctionUsers can self-calibrate power, voltage, and backlight.
LIV Scanning TimeScanning step 0.1~10.0mA configurable,
Step time 2ms~6s optionally configured
PV Scanning TimeScanning step 0.02V~0.1V configurable,
Step time 2ms~6s optionally configured
Pulse Spectrum TestPulse width 200us~650ms,period 2ms~6s,, synchronous output trigger signal
EV Voltage OutputRange:-5.0V ~ +5.0V,Accuracy 0.1% FS±50mV
EV Dark Current TestRange:0 ~ 1mA,Accuracy 0.1% FS±10uA
EV Current TestRange:-200mA ~ +200mA,Accuracy 0.1% FS±2mA
TEC VoltageRange:-1.6V~+1.6V,Accuracy 1% FS±50mV
TEC CurrentRange:-1A~+1A,Accuracy 1% FS±10mA
ThermistorNTC,10K@25℃
Temperature ControlRange:-40℃~+100℃,Accuracy ±0.15℃
Temperature SpeedTypical value 5s (25℃~45℃)

TOP

en_USEnglish